The InH2® Heated Humidifier is designed only for use with specific BMC RESmart® CPAP/Auto CPAP or BPAP devices. Do not use InH2® with any other devices.
The humidifier moistens the air delivered by the BMC RESmart® CPAP/BPAP devices.
The InH2® Heated Humidifier is only used for single patient and must not be re-used on another person. This is to avoid the risk of cross-infection.
The InH2® Heated Humidifier is not intended for use with a patient whose upper airways have been bypassed.
Table of Contents

RESmart®, InH2® are registered trademarks of BMC Medical Co., Ltd.
© BMC Medical Co., Ltd. 2013

Warnings & Cautions

IMPORTANT!
• Read all instructions before using the humidifier.
• Use only with BMC RESmart® devices whose instructions specify the use of this humidifier.
CAUTIONS!
• Indicates the possibility of damage to the device.
• US federal law restricts this device to sale by or on the order of a physician.
• If fluids are spilled onto the humidifier platform, unplug the power cord from the AC wall outlet and allow the humidifier platform to drain and dry before using.
• Take precautions to protect furniture from water damage.
WARNINGS!
• Indicates the possibility for injury to the user or the operator.
• Use the humidifier only for its intended use as described in this manual.
• Use only accessories recommended by BMC.
• Never operate the humidifier if any of the parts are damaged, if it is not working properly, or if the humidifier has been dropped or mishandled. Do not use the humidifier if the water chamber is leaking or damaged in any way. Have any damaged parts replaced before continuing use.
• Never touch the heater plate unless the humidifier is unplugged and the plate has cooled down.
• This equipment is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
• Periodically inspect the power cord for signs of wear or damage. Replace if necessary.
•Do not operate the device in direct sunlight or near a heating appliance because these conditions can increase the temperature of the air coming out of the device.
• When humidifier is used outside the specified ambient temperature range or humidity range, the performance of humidifier will be compromised.

Symbols

 Attention, Consult Accompanying Documents
 Type BF Applied part
 Class II (Double Insulated)
   DC Power
IPX1 Drip-Proof, Vertical
 European CE Declaration of Conformity

Features

Fill Line: This indicates the maximum water level for safe operation.
Outlet Port: Connect the flexible tubing (coaxial 22 mm) here.
Water Chamber: The removable water chamber holds the water for humidification. It has a silica gel cover to outflow the water in cleaning.
Heater Plate: Warms the water in the water chamber.
Indicator Light: When lit, this indicates that the heater plate has been turned on.

Sensor: Do not block this sensor. The humidifier will not operate unless it can detect that the RESmart® device is connected correctly.
Air Inlet with Rubber Connector: Connect to the outlet port on the RESmart® device.
Power Connector: Plug this connector into the power outlet on the BMC RESmart® Device.

Set Up


1. Connect the Humidifier to the RESmart® device. Make sure they are connected completely.
2. Place the RESmart® device and humidifier on a firm, flat surface at a level lower than your sleeping position.
CAUTION!
Do not turn the humidifier on without the water chamber installed.
CAUTION!
Take precautions to protect furniture from water damage.

Daily Use

1. Fill the chamber to the fill line with water (approx. 350 ml) from the outlet port. Distilled water is recommended. Do not overfill the water chamber.
CAUTION!
Always remove the chamber from the humidifier before filling with water.
2. Press down the spring loaded heater plate with the water chamber and slide the chamber into place. Make sure the rubber connector on the inlet port fits securely over the RESmart® device’s air outlet.
3. Connect the flexible tubing to the outlet port on the water chamber.
CAUTION!
Avoid moving or tilting the humidifier when the water chamber has water in it.
4. When the RESmart® device begins blowing, the humidifier will work automatically. The yellow indicator light on the humidifier will turn on. Press the humidifier button can turn off or restart the humidifier.
5. The ideal humidity setting depends on room temperature and humidity. Initially, a setting of 3 is recommended. You can adjust this setting at any time. Please adjust the humidifier setting according to the CPAP User Manual.
IMPORTANT!
When the air flow is turned off, the humidifier will automatically shut off.

Cleaning

The water chamber should be cleaned daily or after each use.
WARNING!
Emptying and cleaning the water chamber daily will help to prevent mold and bacteria growth.
WARNING!
Allow the water in the chamber to cool down to room temperature before removing it from the humidifier.
WARNING!
To avoid electrical shock, disconnect the power cord of RESmart® device before cleaning the humidifier. DO NOT immerse the humidifier into any fluids.
 
Water Chamber
1. Turn the RESmart® device off and allow approximately 15 minutes for the heater plate and water to cool.
2. Disconnect the tubing from the water chamber. Press down on the water chamber and slide it out of the humidifier platform.
3. Open the silica gel cover of water chamber and discard any remaining water. Fill a solution of warm water and a mild dishwashing detergent into the chamber, cover the silica gel cover, rock the chamber a few minutes, and then outflow the solution. Rinse the chamber several times with clean water and allow to air dry.
4. Fill the water chamber and close the silica gel cover. Inspect the water chamber for any leak or damage. Replace the water chamber if any damage is present.
Humidifier Platform
1. Clean the humidifier platform by wiping with a damp cloth. Allow to air dry.
2. Inspect the humidifier platform for any damage and replace if necessary.
Silicone Tube of Humidifier
Clean the Silicone Tube by wiping with an alcohol cotton stick. Allow to air dry.

Service

The humidifier does not require routine servicing.
If the humidifier malfunctions, contact your home care provider immediately. Never attempt to open the humidifier’s enclosure. If necessary, contact your local authorized dealer or BMC Medical Co., Ltd. for technical support and documents.

Specifications

Size
Dimensions:120 × 194 × 112 mm
Weight:< 0.8 kg
Water Capacity:> 350 ml at recommended water level
Product Use, Transport and Storage
                          Operation                               Transport and Storage
Temperature:       5 to 30℃                                -20 to 55℃
Humidity:    ≤ 80% Non-condensing             ≤ 93% Non-condensing
Atmospheric Pressure:  860 to 1060 hPa                  500 to 1060 hPa
Power Requirements
24V DC 1.0A max
Type of Protection Against Electric Shock
Class II Equipment
Degree of Protection Against Electric Shock
Type BF Applied Part
Degree of Protection Against Ingress of Water
IPX1 –Drip-Proof, Vertical
Heater Settings
1 to 5 (104 to 149℉ / 40 to 65℃)
Maximum Operating Pressure
30hPa
Pressure Drop w/ Humidifier
<0.5 hPa at 60 LPM flow
Maximum Delivered Gas Temperature
<40℃
Humidity Range
10 to 45 mg H2O/L

Disposal

When necessary, dispose of the device and accessories in accordance with local regulations.

Traveling With the System

Packing the System
When traveling, the optional CPAP carrying case is for carry-on luggage only. The carrying case will not protect the humidifier if it is put through checked baggage.
Security Stations
For ease at security stations, there is a note on the bottom of the humidifier stating that it is medical equipment. It may be helpful to bring this manual along with you for security personnel.

EMC Requirements

Guidance and manufacturer's declaration - electromagnetic emissions
The RESmart® is intended for use in the electromagnetic environment specified below. The user of the RESmart® should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1,
Class B
The RESmart® uses RF energy only for its internal function. Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Group 1,
Class B
The RESmart® is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Class D
 

Guidance and manufacturer's declaration - electromagnetic immunity
The RESmart® is intended for use in the electromagnetic environment specified below. The user of the RESmart® should make sure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD)
 
IEC 61000-4-2
±6 kV contact
 
 
 
±8 kV air
±6 kV contact
 
 
±8 kV air
Floor should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
 
IEC 61000-4-4
±2 kV for
power supply lines
 
±1 kV for
input/output lines
±2 kV for
power supply lines
 
±1 kV for
input/output lines
Mains power quality should be that of a typical home or hospital.
Surge
 
IEC 61000-4-5
±1 kV
differential mode
 
±2 kV
common mode
±1 kV
differential mode
 
±2 kV
common mode
Mains power quality should be that of a typical home or hospital.
Voltage dips, short interruptions and voltage variations on power supply input lines
 
IEC 61000-4-11
<5 % UT
(>95% dip in UT)
for 0.5 cycle
 
40% UT
(60% dip in UT)
for 5 cycles
 
70% UT
(30% dip in UT)
for 25 cycles
 
<5% UT
(>95% dip in UT)
for 5 s
<5% UT
(>95 % dip in UT)
for 0.5 cycle
 
40% UT
(60% dip in UT)
for 5 cycles
 
70% UT
(30% dip in UT)
for 25 cycles
 
<5% UT
(>95% dip in UT)
for 5 s
Mains power quality should be that of a typical commercial or hospital environment. If the user of the RESmart® requires continued operation during power mains interruptions, it is recommended that the RESmart® be powered from an uninterruptible power supply or from a battery.
Power frequency (50/60 Hz) magnetic field
 
IEC 61000-4-8
3 A/m 0.3 A/m If the pressure deviates more than is indicated in the device specification, it may be necessary to position the RESmart® further from sources of power frequency magnetic fields. The power frequency magnetic field should be measured in the intended installation location to ensure that it is sufficiently low.
NOTE: UT is the AC mains voltage prior to application of the test level.
 
 
 
 
Guidance and manufacturer's declaration - electromagnetic immunity
The RESmart® is intended for use in the electromagnetic environment specified below. The user of the RESmart® should make sure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
 
Conducted RF
IEC 61000-4-6
 
 
Radiated RF
IEC 61000-4-3
 
3 Vrms
150 kHz to 80 MHz
 
 
 
3 V/m
80 MHz to 2.5 GHz
3 Vrms
 
 
 
 
 
3 V/m
Portable and mobile RF communications equipment should be used no closer to any part of the RESmart®, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
 
Recommended separation distance
 
d=1.2√P
d=1.2√P    80 MHz to 800 MHz
d=2.3√P    800 MHz to 2.5 GHz
 
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
 
Field strengths from fixed RF transmitter, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applied.
 
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the RESmart® is used exceeds the applicable RF compliance level above, the RESmart® should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the RESmart®.
 
b Over the frequency range 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
 
 

Warranty

BMC Medical Co., Ltd. warrants that this humidifier shall be free from defects of workmanship and materials and will perform in accordance with the product specifications for a period of one (1) year from the date of sale by BMC Medical Co., Ltd. to the dealer. If the product fails to perform in accordance with the product specifications, BMC Medical Co., Ltd. will repair or replace, at its option, the defective material or part. BMC Medical Co., Ltd. will pay customary freight charges from BMC Medical Co., Ltd. to the dealer location only. This warranty does not cover damage caused by accident, misuse, abuse, alteration and other defects not related to material or workmanship.
To exercise your rights under this warranty, contact your local, authorized dealers or:
 
 
MANUFACTURER:
BMC Medical Co., Ltd.
5/F Main Building, No.19 Gucheng Street West, Shijingshan, Beijing 100043, P.R.China
Tel: 86-10-51663880
Fax: 86-10-51663880 Ext. 810
 
EU AUTHORISED REPRESENTATIVE:
Shanghai International Holding Corp. GmbH (Europe)
Eiffestraβe 80, 20537 Hamburg,Germany
Tel: 0049-40-2513175
Fax: 0049-40-255726世联翻译公司完成医疗器械-加湿器英文翻译

Unitrans世联翻译公司在您身边,离您近的翻译公司,心贴心的专业服务,专业的全球语言翻译与信息解决方案供应商,专业翻译机构品牌。无论在本地,国内还是海外,我们的专业、星级体贴服务,为您的事业加速!世联翻译公司在北京、上海、深圳等国际交往城市设有翻译基地,业务覆盖全国城市。每天有近百万字节的信息和贸易通过世联走向全球!积累了大量政商用户数据,翻译人才库数据,多语种语料库大数据。世联品牌和服务品质已得到政务防务和国际组织、跨国公司和大中型企业等近万用户的认可。 专业翻译公司,北京翻译公司,上海翻译公司,英文翻译,日文翻译,韩语翻译,翻译公司排行榜,翻译公司收费价格表,翻译公司收费标准,翻译公司北京,翻译公司上海。
  • “贵司提交的稿件专业词汇用词准确,语言表达流畅,排版规范, 且服务态度好。在贵司的帮助下,我司的编制周期得以缩短,稿件语言的表达质量得到很大提升”

    华东建筑设计研究总院

  • “我单位是一家总部位于丹麦的高科技企业,和世联翻译第一次接触,心中仍有着一定的犹豫,贵司专业的译员与高水准的服务,得到了国外合作伙伴的认可!”

    世万保制动器(上海)有限公司

  • “我公司是一家荷兰驻华分公司,主要致力于行为学研究软件、仪器和集成系统的开发和销售工作,所需翻译的英文说明书专业性强,翻译难度较大,贵司总能提供优质的服务。”

    诺达思(北京)信息技术有限责任公司

  • “为我司在东南亚地区的业务开拓提供小语种翻译服务中,翻译稿件格式美观整洁,能最大程度的还原原文的样式,同时翻译质量和速度也得到我司的肯定和好评!”

    上海大众

  • “在此之前,我们公司和其他翻译公司有过合作,但是翻译质量实在不敢恭维,所以当我认识刘颖洁以后,对她的专业性和贵公司翻译的质量非常满意,随即签署了长期合作合同。”

    银泰资源股份有限公司

  • “我行自2017年与世联翻译合作,合作过程中十分愉快。特别感谢Jasmine Liu, 态度热情亲切,有耐心,对我行提出的要求落实到位,体现了非常高的专业性。”

    南洋商业银行

  • “与我公司对接的世联翻译客服经理,可以及时对我们的要求进行反馈,也会尽量满足我们临时紧急的文件翻译要求。热情周到的服务给我们留下深刻印象!”

    黑龙江飞鹤乳业有限公司

  • “翻译金融行业文件各式各样版式复杂,试译多家翻译公司,后经过比价、比服务、比质量等流程下来,最终敲定了世联翻译。非常感谢你们提供的优质服务。”

    国金证券股份有限公司

  • “我司所需翻译的资料专业性强,涉及面广,翻译难度大,贵司总能提供优质的服务。在一次业主单位对完工资料质量的抽查中,我司因为俄文翻译质量过关而受到了好评。”

    中辰汇通科技有限责任公司

  • “我司在2014年与贵公司建立合作关系,贵公司的翻译服务质量高、速度快、态度好,赢得了我司各部门的一致好评。贵司经理工作认真踏实,特此致以诚挚的感谢!”

    新华联国际置地(马来西亚)有限公司

  • “我们需要的翻译人员,不论是笔译还是口译,都需要具有很强的专业性,贵公司的德文翻译稿件和现场的同声传译都得到了我公司和合作伙伴的充分肯定。”

    西马远东医疗投资管理有限公司

  • “在这5年中,世联翻译公司人员对工作的认真、负责、热情、周到深深的打动了我。不仅译件质量好,交稿时间及时,还能在我司资金周转紧张时给予体谅。”

    华润万东医疗装备股份有限公司

  • “我公司与世联翻译一直保持着长期合作关系,这家公司报价合理,质量可靠,效率又高。他们翻译的译文发到国外公司,对方也很认可。”

    北京世博达科技发展有限公司

  • “贵公司翻译的译文质量很高,语言表达流畅、排版格式规范、专业术语翻译到位、翻译的速度非常快、后期服务热情。我司翻译了大量的专业文件,经过长久合作,名副其实,值得信赖。”

    北京塞特雷特科技有限公司

  • “针对我们农业科研论文写作要求,尽量寻找专业对口的专家为我提供翻译服务,最后又按照学术期刊的要求,提供润色原稿和相关的证明文件。非常感谢世联翻译公司!”

    中国农科院

  • “世联的客服经理态度热情亲切,对我们提出的要求都落实到位,回答我们的问题也非常有耐心。译员十分专业,工作尽职尽责,获得与其共事的公司总部同事们的一致高度认可。”

    格莱姆公司

  • “我公司与马来西亚政府有相关业务往来,急需翻译项目报备材料。在经过对各个翻译公司的服务水平和质量的权衡下,我们选择了世联翻译公司。翻译很成功,公司领导非常满意。”

    北京韬盛科技发展有限公司

  • “客服经理能一贯热情负责的完成每一次翻译工作的组织及沟通。为客户与译员之间搭起顺畅的沟通桥梁。能协助我方建立专业词库,并向译员准确传达落实,准确及高效的完成统一风格。”

    HEURTEY PETROCHEM法国赫锑石化

  • “贵公司与我社对翻译项目进行了几次详细的会谈,期间公司负责人和廖小姐还亲自来我社拜访,对待工作热情,专业度高,我们双方达成了很好的共识。对贵公司的服务给予好评!”

    东华大学出版社

  • “非常感谢世联翻译!我们对此次缅甸语访谈翻译项目非常满意,世联在充分了解我司项目的翻译意图情况下,即高效又保质地完成了译文。”

    上海奥美广告有限公司

  • “在合作过程中,世联翻译保质、保量、及时的完成我们交给的翻译工作。客户经理工作积极,服务热情、周到,能全面的了解客户的需求,在此表示特别的感谢。”

    北京中唐电工程咨询有限公司

  • “我们通过图书翻译项目与你们相识乃至建立友谊,你们报价合理、服务细致、翻译质量可靠。请允许我们借此机会向你们表示衷心的感谢!”

    山东教育出版社

  • “很满意世联的翻译质量,交稿准时,中英互译都比较好,措辞和句式结构都比较地道,译文忠实于原文。TNC是一家国际环保组织,发给我们美国总部的同事后,他们反应也不错。”

    TNC大自然保护协会

  • “原英国首相布莱尔来访,需要非常专业的同声传译服务,因是第一次接触,心中仍有着一定的犹豫,但是贵司专业的译员与高水准的服务,给我们留下了非常深刻的印象。”

    北京师范大学壹基金公益研究院

  • “在与世联翻译合作期间,世联秉承着“上善若水、厚德载物”的文化理念,以上乘的品质和质量,信守对客户的承诺,出色地完成了我公司交予的翻译工作。”

    国科创新(北京)信息咨询中心

  • “由于项目要求时间相当紧凑,所以世联在保证质量的前提下,尽力按照时间完成任务。使我们在世博会俄罗斯馆日活动中准备充足,并受到一致好评。”

    北京华国之窗咨询有限公司

  • “贵公司针对客户需要,挑选优秀的译员承接项目,翻译过程客户随时查看中途稿,并且与客户沟通术语方面的知识,能够更准确的了解到客户的需求,确保稿件高质量。”

    日工建机(北京)国际进出口有限公司

15801211926

18017395793
点击添加微信

无需转接等回电